-
Notifications
You must be signed in to change notification settings - Fork 0
Description
Last week on 9/7/22, the bulb of the titrator was left out accidentally for about 2 hours. I attempted the pH calibration which measured values within the accepted range but measured about 9mV as opposed to 0 for the pH 7 buffer. I also ran a CRM but this deviated from the accepted value by over +1%. I chose not to continue with titrations and let the electrode sit in KCl for one week.
Today, Jill and I attempted the pH calibration and noticed that the values for pH 4, 7, and 10 were drifting rather than oscillating between two points and settling on a value.
We flushed the internal and external KCl reservoirs twice and refilled the reservoirs with new solution. We let the probe soak in pH 4 buffer for about 4 hours to reactivate the electrode. After this, we calibrated the pH probe again with new buffers and all values fell within the accepted range according to the manufacturers reference guide. We ran a CRM with a different bottle of CRM that had been previously tested, leading us to believe that the CRM may have not been accurate.
While we resolved that we do not need to replace the probe, we should look out for values drifting during the calibration or TA measurements. The probe is 1-2 years old, so we may need a replacement in the near future (within 6 months- 1 year).
On 9/30/22, Jill, Amy and I replaced the acid titrant and completed a run with two new bottles of CRM. Here are the values we recorded that indicate probe deprecation. For the pH calibration and the CRM values, deviance from the expected was first observed the day the probe was left out (8/25/22). Since 8/25/22, pH calibration values have trended higher and higher for ph 4 and 7 and less negative for pH 10. The deviance is progressively more dramatic with each sample ie: deviance from expected at pH 10 (Sample #3) is more than the deviance at pH 4.
We also tested two CRM bottles, one I had tested (Batch # 176, Bottle A) and a new bottle from the Putnam batch bottled on March 18, 2022 (Batch #202) and also ran a junk sample. The CRM values were within range today, but have deviated from the known value by over 1% for the past few weeks.
slope: -57.06
zero pt: 7.289
pH 4: 187.2
pH 7: 16.4
pH 10: -157.2
CRM176: -0.299%
CRM202: -.223 %
We were not able to calculate TA values because the pH calibrations were so far from the expected values that the R script failed, so we used the pH calibration from last week.
"Error in nls(y ~ ((At + (St/(1+ Ks * Z/(f * Hprime))) + (Ft/(1 + Kf/f * : number of iterations exceeded maximum of 50". We researched this error and it typically occurs when the model does not fit your data.